1. General Description
E2 Class Calibration Weights are high-precision mass standards used for calibration of analytical balances and for verification/calibration of F1 class weights. These weights provide very low uncertainty and are intended for use in controlled laboratory environments.
2. Standards & Metrological Compliance
-
Accuracy Class: E2
-
International Standard: OIML R-111-1
-
Calibration Standard: ISO/IEC 17025
-
Traceability:
-
Traceable to National Physical Laboratory (NPL), India or equivalent national standard
-
SI unit of mass (kilogram)
-
-
Calibration Requirement: NABL-accredited calibration recommended / mandatory for laboratory use
3. Nominal Weight Range
-
1 mg to 500 mg (wire / foil type)
-
1 g to 20 kg (solid cylindrical type)
4. Metrological Characteristics (Typical)
| Nominal Mass | Maximum Permissible Error (±) | Expanded Uncertainty U (k=2) |
|---|---|---|
| 20 kg | 1.0 mg | ≤ 0.33 mg |
| 10 kg | 0.5 mg | ≤ 0.17 mg |
| 5 kg | 0.25 mg | ≤ 0.08 mg |
| 2 kg | 0.10 mg | ≤ 0.03 mg |
| 1 kg | 0.05 mg | ≤ 0.017 mg |
| 100 g | 0.015 mg | ≤ 0.005 mg |
| 10 g | 0.003 mg | ≤ 0.001 mg |
| 1 g | 0.001 mg | ≤ 0.0003 mg |
(As per OIML R-111; actual values stated in NABL certificate)
5. Material Specifications
-
Material: High-purity non-magnetic austenitic stainless steel
-
Grade: Special metrology-grade SS (vacuum melted)
-
Density: (8000 ± 20) kg/m³
-
Magnetic Susceptibility: ≤ 0.02
-
Residual Magnetism: ≤ 1 µT
-
Surface Reactivity: Low adsorption, corrosion resistant
6. Construction & Surface Finish
-
Design:
-
Cylindrical with adjusting cavity (g to kg)
-
Foil / wire type (mg range)
-
-
Manufacturing: Precision CNC machining
-
Surface Finish:
-
Mirror polished
-
Surface roughness Ra ≤ 0.2 µm
-
-
Edges: Rounded to minimize air turbulence effects
7. Dimensional & Thermal Stability
-
Coefficient of Thermal Expansion:
~16 × 10⁻⁶ /°C -
Long-Term Stability:
< 1/10 of MPE per year under controlled storage -
Mass Drift: Minimal when stored in protective case
8. Air Buoyancy & Density Correction
-
Air buoyancy correction applied during calibration
-
Individual density value stated in calibration certificate
-
Suitable for mass comparator-based calibration
9. Environmental Reference Conditions (Calibration)
-
Temperature: 20 °C ± 1 °C
-
Relative Humidity: 40 % – 60 % RH
-
Atmospheric Pressure: 950 – 1050 hPa
-
Vibration: Negligible
10. Handling & Usage Requirements
-
Handle only with gloves or tweezers
-
Avoid direct hand contact
-
Allow thermal stabilization before use
-
Use in draft-free, temperature-controlled lab
-
Clean with lint-free cloth or approved cleaning method
11. Storage & Packaging
-
Individual cavities in wooden / aluminium / ABS case
-
Dust-free, shock-resistant storage
-
Anti-corrosion lining (optional)
12. NABL Calibration Certificate Includes
-
NABL logo & accreditation number
-
Nominal & conventional mass values
-
Expanded uncertainty (k=2)
-
Density of weight
-
Environmental conditions
-
Traceability statement
-
Calibration & due date
13. Applications
-
Calibration of analytical balances
-
Calibration of F1 class weights
-
NABL / ISO-17025 laboratories
-
Pharmaceutical & chemical laboratories
-
Research & academic institutions
14. Compliance Summary
✔ OIML R-111 compliant
✔ NABL / ISO-17025 traceable
✔ High stability & low uncertainty
✔ Suitable for laboratory & audit use